Comparison of high-performance liquid chromatography/tandem mass spectrometry and high-performance liquid chromatography/photo-diode array detection for the quantitation of carotenoids, retinyl esters, α-tocopherol and phylloquinone in chylomicron-rich fractions of human plasma

高效液相色谱/串联质谱法和高效液相色谱/光电二极管阵列检测法对富含乳糜微粒的人体血浆中类胡萝卜素、视黄酯、α-生育酚和叶绿醌的定量分析比较

阅读:5
作者:Rachel E Kopec, Ralf M Schweiggert, Ken M Riedl, Reinhold Carle, Steven J Schwartz

Conclusions

HPLC/MS/MS was more sensitive than HPLC-PDA for six of the eight analytes and represents a powerful tool for the analysis of chylomicron samples and potentially other biological samples of limited sample size. When internal standards are available for the target carotenoid, employing MS/MS detection may reduce the necessary blood sample volume, which is particularly advantageous for minimizing risk and discomfort to human subjects during clinical studies.

Methods

After selecting an efficient extraction method for the analytes, both the HPLC/PDA and the HPLC/MS/MS methods were developed and several parameters validated using an HP 1200 series HPLC system interfaced with a HP 1200 series diode-array detector (Agilent Technologies, Santa Clara, CA, USA) and a QTRAP 5500 (AB Sciex, Foster City, CA, USA) via an atmospheric pressure chemical ionization (APCI) probe operated in positive ion mode.

Results

For lycopene, α- and β-carotene, HPLC/MS/MS was up to 37 times more sensitive than HPLC-PDA. PDA detection was shown to be up to 8 times more sensitive for lutein. MS/MS signals were enhanced by matrix components for lutein and β-cryptoxanthin, as determined by referencing to the matrix-independent PDA signal. In contrast, matrix suppression was observed for retinyl palmitate, α-carotene, and β-carotene. Both detectors showed similar suitability for α-tocopherol, lycopene and retinyl palmitate (representing ~73% of total retinyl esters). MS/MS exclusively allowed the quantitation of minor retinyl esters, phylloquinone, and (Z)-lycopene isomers. Conclusions: HPLC/MS/MS was more sensitive than HPLC-PDA for six of the eight analytes and represents a powerful tool for the analysis of chylomicron samples and potentially other biological samples of limited sample size. When internal standards are available for the target carotenoid, employing MS/MS detection may reduce the necessary blood sample volume, which is particularly advantageous for minimizing risk and discomfort to human subjects during clinical studies.