Downsloping TP Segment in the Precordial Leads on a Standard 12-lead Electrocardiogram: Suspected Cardiac Impulse-tapping Artifact

标准12导联心电图胸前导联TP段下斜:疑似心脏冲动敲击伪迹

阅读:1

Abstract

We herein report an 80-year-old man showing a downsloping TP segment together with an increase in the height of the T wave in the precordial leads on a standard 12-lead electrocardiogram (ECG). Separately, an 87-year-old woman showed only a downsloping TP segment in the precordial leads on a standard 12-lead ECG. Neither patient reported chest pain or dyspnea when ECGs was obtained. This downsloping TP segment in the precordial leads on the standard 12-lead ECG is thought to be due to a cardiac impulse-tapping artifact. Differential diagnoses are also discussed.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。