Safety of Magnetic Resonance Imaging in Patients with Cardiac Implantable Electronic Devices

心脏植入式电子设备患者磁共振成像的安全性

阅读:1

Abstract

BACKGROUND: MRI (magnetic resonance imaging) represents the diagnostic image modality of choice in several conditions. With an increasing number of patients requiring MRI for diagnostic purposes, the issue of safety in patients with cardiac implantable electronic devices (CIED) undergoing this imaging modality will play an ever more important role. The purpose of this study was to assess the safety and device function following MRI in an unrestricted real-world cohort of patients with a wide array of cardiac devices. METHODS: We conducted a retrospective single-center study including 1010 MRI studies conducted in adult patients (≥18 years) with an implanted CIED treated in the University Hospital of Munich (LMU) between July 2012 and March 2024. Patients with non-MR conditionally labeled leads, abandoned or epicardial leads, as well as lead fragments, were included for analysis. RESULTS: Across a total of 1010 MRIs (920 total MR-conditional device generators) performed in patients with an implanted CIED, there were no deaths, reports of discomfort, palpitations, heating, or ventricular arrythmias in the 24 h following MRI. Only 2/1010 MRIs were followed by a reported atrial arrhythmia within 24 h, both in patients with an MR-conditional pacemaker (PM) device without an abandoned lead. No significant changes in device function following MRI from baseline were observed across all included CIEDs. Lastly, no instances of severe malfunction, such as generator failure, loss of capture, electrical reset, or inappropriate inhibition of pacing, were found in post-MRI interrogation reports across all MRI studies. CONCLUSIONS: Based on the analysis of 1010 MRIs undergone by patients with CIEDs, following standardized device interrogation, manufacturer-advised device programming, monitoring of vital function, and manufacturer-advised reprogramming, MRI can be performed safely and without adverse events or changes in device function.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。