Sudden Occurrence of Pacemaker Capture Failure during Irreversible Electroporation Ablation for Prostate Cancer in Post-COVID-19 Patient: A Case Report

新冠肺炎后患者行前列腺癌不可逆电穿孔消融术期间突发起搏器夺获失败:病例报告

阅读:2

Abstract

Irreversible electroporation (IRE) ablation is a novel treatment option for localized prostate cancer. Here, we present a case of an abrupt and fatal arrhythmia during the IRE procedure in a prostate cancer patient with an implanted permanent pacemaker. A 78-year-old male patient with a pacemaker due to sick sinus syndrome and syncope was scheduled for IRE prostate ablation surgery under general anesthesia. He had a history of recovering from coronavirus disease 2019 (COVID-19) after having been vaccinated against it and recovered without sequalae. Pacemaker interrogation and reprogramming to asynchronous AOO mode was carried out before surgery, however, sinus pause occurred repeatedly during ablation pulse delivery. After the first sinus pause of 2.25 s there was a decrease in continuous arterial blood pressure (ABP). During the delivery of the second and third pulses, identical sinus pauses were observed due to failure to capture. However, the atrial-paced rhythm recovered instantly, and vital signs became acceptable. Although sinus pause recovered gradually, the duration thereof was increased by the delivery of more IRE pulses, with a subsequent abrupt decrease seen in blood pressure. The pacemaker was urgently reprogrammed to DOO mode, after which there were no further pacing failures and no hemodynamic adverse events. For patients with pacemakers, close cardiac monitoring in addition to the interrogation of the pacemaker during the electromagnetic interference (EMI) procedure is recommended, especially in the case of having a disease that may aggravate cardiac vulnerability, such as COVID-19.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。