Influence of the arterial input sampling location on the diagnostic accuracy of cardiovascular magnetic resonance stress myocardial perfusion quantification

动脉输入采样位置对心血管磁共振应激心肌灌注定量诊断准确性的影响

阅读:1

Abstract

BACKGROUND: Quantification of myocardial blood flow (MBF) and myocardial perfusion reserve (MPR) by cardiovascular magnetic resonance (CMR) perfusion requires sampling of the arterial input function (AIF). While variation in the AIF sampling location is known to impact quantification by CMR and positron emission tomography (PET) perfusion, there is no evidence to support the use of a specific location based on their diagnostic accuracy in the detection of coronary artery disease (CAD). This study aimed to evaluate the accuracy of stress MBF and MPR for different AIF sampling locations for the detection of abnormal myocardial perfusion with expert visual assessment as the reference. METHODS: Twenty-five patients with suspected or known CAD underwent vasodilator stress-rest perfusion with a dual-sequence technique at 3T. A low-resolution slice was acquired in 3-chamber view to allow AIF sampling at five different locations: left atrium (LA), basal left ventricle (bLV), mid left ventricle (mLV), apical left ventricle (aLV) and aortic root (AoR). MBF and MPR were estimated at the segmental level using Fermi function-constrained deconvolution. Segments were scored as having normal or abnormal perfusion by visual assessment and the diagnostic accuracy of stress MBF and MPR for each location was evaluated using receiver operating characteristic curve analysis. RESULTS: In both normal (300 out of 400, 75 %) and abnormal segments, rest MBF, stress MBF and MPR were significantly different across AIF sampling locations (p < 0.001). Stress MBF for the AoR (normal: 2.42 (2.15-2.84) mL/g/min; abnormal: 1.71 (1.28-1.98) mL/g/min) had the highest diagnostic accuracy (sensitivity 80 %, specificity 85 %, area under the curve 0.90; p < 0.001 versus stress MBF for all other locations including bLV: normal: 2.78 (2.39-3.14) mL/g/min; abnormal: 2.22 (1.83-2.48) mL/g/min; sensitivity 91 %, specificity 63 %, area under the curve 0.81) and performed better than MPR for the LV locations (p < 0.01). MPR for the AoR (normal: 2.43 (1.95-3.14); abnormal: 1.58 (1.34-1.90)) was not superior to MPR for the bLV (normal: 2.59 (2.04-3.20); abnormal: 1.69 (1.36-2.14); p = 0.717). CONCLUSIONS: The AIF sampling location has a significant impact on MBF and MPR estimates by CMR perfusion, with AoR-based stress MBF comparing favorably to that for the current clinical reference bLV.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。