[Research of the training methods and structure design of intelligent knee prosthesis based on physiological gait]

【基于生理步态的智能膝关节假肢训练方法及结构设计研究】

阅读:1

Abstract

The performance of intelligent prosthetic knee has an important effect on the realization of physiological gait of transfemoral amputees. A new type of single axis hydraulic damping knee prosthesis was designed based on the analysis of physiological gait. The training methods of the stance and swing phase were proposed. Knee prosthesis test was done through simulation and measurement device. The control target of peak flexion angle during swing of knee prosthesis is chosen to be 60-70°. When the damper valve closure was 0%, maximum swing-phase knee flexion angle of knee prosthesis were (86±2)°, (91±3)° and (97±3)° with the speed of 0.8 m/s, 1.2 m/s and 1.8 m/s, respectively. Once the valve closure was changed, maximum swing-phase knee flexion angle with different speeds could be adjusted between 60° and 70° and the required valve closure percentage were separately 25%, 40% and 70%. The damping adjustment law of intelligent knee prosthesis to achieve physiological gait was revealed.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。