Gait asymmetry of transfemoral amputees using mechanical and microprocessor-controlled prosthetic knees

使用机械式和微处理器控制式假肢膝关节的股骨截肢患者的步态不对称性

阅读:3

Abstract

BACKGROUND: Amputees walk with an asymmetrical gait, which may lead to future musculoskeletal degenerative changes. The purpose of this study was to compare the gait asymmetry of active transfemoral amputees while using a passive mechanical knee joint or a microprocessor-controlled knee joint. METHODS: Objective 3D gait measurements were obtained in 15 subjects (12 men and 3 women; age 42, range 26-57). Research participants were longtime users of a mechanical prosthesis (mean 20 years, range 3-36 years). Joint symmetry was calculated using a novel method that includes the entire waveform throughout the gait cycle. FINDINGS: There was no significant difference in hip, knee and ankle kinematics symmetry when using the different knee prostheses. In contrast, the results demonstrated a significant improvement in lower extremity joint kinetics symmetry when using the microprocessor-controlled knee. INTERPRETATION: Use of the microprocessor-controlled knee joint resulted in improved gait symmetry. These improvements may lead to a reduction in the degenerative musculoskeletal changes often experienced by amputees.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。