Visual Evaluation of Image Quality of a Low Dose 2D/3D Slot Scanner Imaging System Compared to Two Conventional Digital Radiography X-ray Imaging Systems

低剂量二维/三维狭缝扫描成像系统与两种传统数字放射成像X射线成像系统的图像质量视觉评价

阅读:1

Abstract

The purpose of this study was to assess the image quality of the low dose 2D/3D slot scanner (LDSS) imaging system compared to conventional digital radiography (DR) imaging systems. Visual image quality was assessed using the visual grading analysis (VGA) method. This method is a subjective approach that uses a human observer to evaluate and optimise radiographic images for different imaging technologies. METHODS AND MATERIALS: ten posterior-anterior (PA) and ten lateral (LAT) images of a chest anthropomorphic phantoms and a knee phantom were acquired by an LDSS imaging system and two conventional DR imaging systems. The images were shown in random order to three (chest) radiologists and three experienced (knee) radiographers, who scored the images against a number of criteria. Inter- and intraobserver agreement was assessed using Fleiss' kappa and weighted kappa. RESULTS: the statistical comparison of the agreement between the observers showed good interobserver agreement, with Fleiss' kappa coefficients of 0.27-0.63 and 0.23-0.45 for the chest and knee protocols, respectively. Comparison of intraobserver agreement also showed good agreement with weighted kappa coefficients of 0.27-0.63 and 0.23-0.45 for the chest and knee protocols, respectively. The LDSS imaging system achieved significantly higher VGA image quality compared to the DR imaging systems in the AP and LAT chest protocols (p < 0.001). However, the LDSS imaging system achieved lower image quality than one DR system (p ≤ 0.016) and equivalent image quality to the other DR systems (p ≤ 0.27) in the knee protocol. The LDSS imaging system achieved effective dose savings of 33-52% for the chest protocol and 30-35% for the knee protocol compared with DR systems. CONCLUSIONS: this work has shown that the LDSS imaging system has the potential to acquire chest and knee images at diagnostic quality and at a lower effective dose than DR systems.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。