Survival rate of cemented versus cementless tibial component in primary total knee arthroplasty over 5 years of follow-up: comparative study of 109 prostheses

初次全膝关节置换术中,有骨水泥型与无骨水泥型胫骨假体5年随访生存率比较:109例假体的对比研究

阅读:1

Abstract

INTRODUCTION: Knee osteoarthritis is the main indication for primary total knee arthroplasty (TKA). It is now accepted that cementless implantation of the femoral component provides equivalent results to cemented one, however, the optimal fixation method of the tibial component remains controversial. The purpose of this study was to compare the survivorship of cemented versus cementless tibial baseplate in primary total knee arthroplasty. MATERIALS AND METHODS: We carried out a retrospective, monocentric study, including 109 TKA (Zimmer(®) Natural Knee II ultra-congruent mobile-bearing) implanted between 2004 and 2010 for primary osteoarthritis, comparing 2 groups depending on tibial component fixation method, one cemented (n = 68) and one cementless (n = 41). Clinical (Knee Society Rating System (KSS), Hospital for Special Surgery (HSS) scores, range of motion) and radiodiological outcomes were assessed at last follow-up with a minimal follow-up of 5 years. RESULTS: Mean follow-up were 8.14 [5.31-12.7] and 8.06 [5.22-12.02] years, respectively, in cemented and cementless groups. The tibial component survival rate was 100% [95CI: 91.4-100] in the cementless group and 97.1% [95CI: 89.78-99.42] in the cemented group (2 aseptic loosenings) (p = 0.27). Radiolucent lines were present in 31.7% (n = 13) of the cementless and 44.1% (n = 30) of the cemented baseplates (p = 0.2). The postoperative KSS knee score was higher in the cementless group (99 ± 3 vs. 97 ± 7.5; p = 0.02), but there was no significant difference in KSS function, global KSS and HSS scores. Mean range of flexion was 120 ± 10° in the cementless group and 122.5 ± 15° in the cemented group (p = 0.37). No significant differences were found on the radiographic data or on complications. CONCLUSION: In this study, the survival rate of the tibial component is not influenced by its fixation method at a mean follow-up of 8 years in osteoarthritis, which confirms the reliability of cementless fixation in primary TKA.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。