A Novel In Vivo Joint Loading System to Investigate the Effect of Daily Mechanical Load on a Healing Anterior Cruciate Ligament Reconstruction

一种新型体内关节负荷系统,用于研究日常机械负荷对前交叉韧带重建术后愈合的影响

阅读:1

Abstract

We designed and validated a novel knee joint fixation/distraction system to study tendon-to-bone healing in an in vivo rat model of anterior cruciate ligament (ACL) reconstruction. The system uses an external fixator to apply a cyclic distraction of the knee joint while monitoring the resultant force developed across the joint, thus providing a temporal indication of structural changes during the healing process of the bone-tendon-bone reconstruction. The validation was performed using an optical kinematic tracking system to determine the local displacement of the knee. The average system compliance was determined to be 42.4 +/- 8.8 mum/N with a coefficient of variation of 20.7%. The compliance was used to obtain a best fit correction factor which brought the total root mean square error of knee joint distraction to within 179 mum (16.1%) of the applied distraction. We performed a pilot study using 15 rats that had ACL reconstructions using a flexor digitorum longus tendon autograft and found that the animals tolerated the indwelling fixator and daily anesthesia over a 10 day loading protocol. Our knee joint fixation/distraction system provides a valuable tool to study how mechanical stimuli affect in vivo bone-tendon-bone healing.