Delayed migration of a Sapien 3 Ultra Resilia following transcatheter aortic valve implantation after selection of a smaller-sized valve

经导管主动脉瓣置换术后,由于选择了较小尺寸的瓣膜,Sapien 3 Ultra Resilia 瓣膜发生延迟移位

阅读:2

Abstract

This report discusses a rare case of delayed migration of a Sapien 3 Ultra Resilia (S3UR) valve following transcatheter aortic valve implantation. An 81-year-old Japanese woman had a borderline aortic annular size of 20-23 mm according to the manufacturer's size chart. We chose to implant a smaller S3UR of 20 mm with an 80/20 depth ratio to allow for a second intervention, ensuring good hemodynamics and minimizing paravalvular leak. The patient initially had a favorable outcome despite an accidental 50/50 depth ratio during implantation. On postoperative day 3, the S3UR migrated into the left ventricular outflow tract. Emergency surgical aortic valve replacement was performed to retrieve the migrated valve. Use of the S3UR has led to a growing preference for smaller valve sizes. However, the risk of migration should be recognized. When an accidental 50/50 depth ratio implantation is encountered, post-dilation or second valve implantation should be performed immediately.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。