A snare maneuver to reposition self-expandable aortic valve after transcatheter replacement of surgically implanted valve

经导管置换外科植入式主动脉瓣膜后,采用圈套器操作重新定位自膨式主动脉瓣膜

阅读:1

Abstract

In this case, we successfully repositioned the transcatheter heart valve (THV) by pulling it with a snare inserted via the right brachial artery. Attempting to pull the THV via the femoral approach was not successful, due to the vector of the pulling force not being coaxial. Changing the direction of the force by switching to an approach from the right brachial artery successfully prevented perivalvular leakage. While there are previous case reports of repositioning an implanted transcatheter valve in a native valve using a gooseneck snare, to our knowledge, this is the first case of successfully repositioning an implanted transcatheter valve in a valve-in-valve procedure using the precise assessment that can be achieved with intra-procedural transesophageal echocardiography. LEARNING OBJECTIVE: Perivalvular leakage (PVL) is a major problem after transcatheter aortic valve replacement. In this case, we experienced PVL after implantation of self-expandable transcatheter heart valve (THV) inside the previously inserted Mosaic bioprosthetic valve. Depth of the implanted valve was too low into left ventricle accessed with transesophageal echocardiogram. It is barely known whether THV in surgical implanted aortic valve could be repositioned using snare maneuver, but we successfully pulled the valve via brachial artery and PVL was controlled.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。