Experiences from using LC-MS/MS for analysis of immunosuppressive drugs in a TDM service

在 TDM 服务中使用 LC-MS/MS 分析免疫抑制药物的经验

阅读:6
作者:Anton Pohanka, Staffan Rosenborg, Jonatan D Lindh, Olof Beck

Abstract

A liquid chromatography-tandem mass spectrometry method is presented that was applied for over five years for routine measurement of the immunosuppressant drugs ciclosporin, everolimus, sirolimus and tacrolimus in blood. The method has been used for analysis of 142 thousand unknowns and has been running 7days a week without a single day shutdown during the entire time. The measuring ranges were 10-1500ng/mL for ciclosporin and 1-50ng/mL for everolimus, sirolimus and tacrolimus. The method validation showed performance meeting the EMA validation guideline requirements and acceptable performance in a proficiency control program. The routine work was based on daily calibrations on two parallel instrument systems, one with multiplexing providing a capacity of 55 injections/h. During week-days up to six batches of samples were run to meet requirements of short reporting times. Although there is still room for method improvements our main conclusion is that the liquid chromatography-tandem mass spectrometry technique provides a viable platform for an analytical routine service for therapeutic drug monitoring of immunosuppressant drugs that meet clinical needs of reporting times and reliable results. We also describe early involvement and education of clinicians to gain acceptance and correct clinical interpretation of analytical results more accurate and precise, but differing from those obtained by immunoassay.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。