Reliability of measuring the fat content of the lumbar vertebral marrow and paraspinal muscles using MRI mDIXON-Quant sequence

使用 MRI mDIXON-Quant 序列测量腰椎骨髓和椎旁肌脂肪含量的可靠性

阅读:1

Abstract

PURPOSE: We aimed to assess the reliability of measuring the fat content of the lumbar vertebral marrow and the paraspinal muscles using magnetic resonance imaging (MRI) mDIXON-Quant sequence. METHODS: Thirty-one healthy volunteers were included. All participants underwent liver mDIXON-Quant imaging on a 3.0 T Philips MRI scanner by observer A. Within two weeks, observer B repeated the scan. After the examination, each observer independently measured the fat content of the third lumbar vertebra (L3), and the psoas (PS), erector spinae (ES), and multifidus (MF) muscles on central L3 axial images. After two weeks, each observer repeated the same measurements. They were blinded to their previous results. Reliability was estimated by evaluating the repeatability and reproducibility. RESULTS: The repeatability of the fat content measurements of L3, PS, ES, and MF was high. The intraclass correlation coefficients of the fat content of L3, PS, ES, and MF were 0.997, 0.984, 0.997, and 0.995 for observer A and 0.948, 0.974, 0.963, and 0.995 for observer B, respectively. The reproducibility of the measurement of the fat content of L3, PS, ES, and MF was high, and the interclass correlation coefficients were 0.984, 0.981, 0.977, and 0.998, respectively. CONCLUSION: Using mDIXON-Quant imaging to measure the fat content of the lumbar vertebral marrow and paraspinal muscles shows high reliability and is suitable for use in clinical practice.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。