Capillary and Venous Blood Glucose Accuracy in Blood Glucose Meters Versus Reference Standards: The Impact of Study Design on Accuracy Evaluations

血糖仪与参考标准在毛细血管和静脉血糖测量准确性方面的比较:研究设计对准确性评估的影响

阅读:1

Abstract

BACKGROUND: Anecdotal blood glucose assessments conducted by health care professionals (HCPs) in the field have highlighted differences in results when methodology used is not according to best practices for measuring blood glucose. This study assessed the impact on accuracy of blood glucose measurements when methodology deviates from the recommended study design and recommended reference instrument. METHODS: Adults with type 1 or type 2 diabetes provided capillary and venous blood samples for accuracy assessments using OneTouch® Verio® (Verio) and OneTouch® Ultra 2® (Ultra) blood glucose meters (BGM) and two different reference instruments. RESULTS: Increases in mean bias were observed when comparing capillary to venous samples tested on the BGMs and the recommended reference instrument. Mean bias was even greater when a hospital blood glucose analyzer was used to measure venous plasma glucose. Increases in mean bias observed for Ultra BGM when testing venous blood on the meter compared to the recommended reference instrument was likely due to the interfering effects of low oxygen levels in the venous blood sample. Conversely, Verio meters, which are insensitive to low oxygen levels, showed little difference from baseline when testing venous blood on the meter compared to results from the same venous sample measured on a reference instrument. CONCLUSIONS: Deviations from the best practice study design of comparing capillary blood glucose results tested on the blood glucose meter with the manufacturer's stated reference instrument will affect accuracy of blood glucose measurements.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。