Testing a new surfactant in a widely-used blood mimic for ultrasound flow imaging

在广泛使用的血液模拟物中测试一种新型表面活性剂,用于超声流动成像

阅读:1

Abstract

BACKGROUND: A blood-mimicking fluid developed by Ramnarine et al. has been widely used in flow phantoms for ultrasound flow imaging research, and it has also been cited by IEC 61685 as a reference for making blood-mimicking fluid.However, the surfactant material Synperonic N in this blood-mimicking fluid recipe is phased out from the European market due to environmental issues. The aim of this study is to test whether Synperonic N can be substituted by biodegradable Synperonic A7 in making blood-mimicking fluid for ultrasound flow imaging research. METHODS AND MATERIALS: A flow phantom was fabricated to test the blood-mimicking fluid with Synperonic N and Synperonic A7 as surfactants separately. Doppler images and velocity data were collected using a clinical ultrasound scanner under constant and pulsatile flows; and images and measured velocities were compared. RESULTS: It was found that both blood mimics can provide exactly the same images under spectral Doppler ultrasound and colour Doppler ultrasound in terms of their image qualities. The maximum velocities under constant flow were measured by the spectral Doppler ultrasound as 0.4714 ± 0.001 m.s(-1) and 0.4644 ± 0.001 m.s(-1) for blood-mimicking fluid with Synperonic N and blood-mimicking fluid with Synperonic A7, respectively. Measured velocities using the two different blood-mimicking fluids were statistically different (p < 0.001), but this difference was less than 2%. The Synperonic A7 can be used as a substitute for Synperonic N as a surfactant material in making the blood-mimicking fluid for ultrasound flow imaging research.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。