Higher blood glucose and larger fluctuations detected postoperatively using continuous glucose monitoring: a preliminary study following total knee or hip arthroplasty

术后采用连续血糖监测发现血糖升高和波动幅度增大:一项针对全膝关节或髋关节置换术的初步研究

阅读:1

Abstract

BACKGROUND: The control of diabetes mellitus (DM) should help reduce the incidence of periprosthetic joint infection (PJI). Self-monitoring of blood glucose (SMBG) concentration is usually undertaken at fixed time-points. Therefore, the extent of postoperative blood glucose fluctuation might be underestimated. To provide a more comprehensive assessment, continuous glucose monitoring (CGM) is beginning to be used. However, no previous studies have evaluated blood glucose concentrations using CGM following orthopedic surgery. Therefore, the differences between the maximum blood glucose concentrations measured using SMBG and CGM, and the mean amplitude of the glycemic fluctuation in patients with frank diabetes mellitus (DM) or pre-diabetes were evaluated. Blood glucose was measured in 20 patients who had undergone total hip or total knee arthroplasty (12 patients with DM and eight with pre-diabetes). Patients were fitted with a CGM device in the operating room, which was worn for 6 days postoperatively, and used to evaluate blood glucose concentration continuously. SMBG was performed simultaneously for the same period. RESULTS: The mean difference between the maximum blood glucose concentrations measured using SMBG and CGM was 25.0 ± 20.3 mg/dl (range, - 17 to 81 mg/dl), with the concentrations measured using CGM tending to be higher than those measured using SMBG (P = 0.04). Blood glucose concentrations measured using CGM tended to be higher than those measured using SMBG until postoperative day 2, and to decrease gradually after postoperative day 4. There were no significant differences in the standard deviation of the blood glucose concentrations between the two groups. CONCLUSIONS: Blood glucose concentrations > 200 mg/dl and larger fluctuations were more frequently recorded using CGM than SMBG, especially until postoperative day 2. Thus, CGM is more useful for the identification of high blood glucose concentrations and larger fluctuations. However, this information was not provided in real time.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。