Determination of reference intervals and comparison of venous blood gas parameters using standard and non-standard collection methods in 24 cats

采用标准和非标准采集方法,对24只猫的静脉血气参数进行参考区间测定和比较。

阅读:1

Abstract

Objectives The aim of this study was to determine in-house reference intervals (RIs) for venous blood analysis with the RAPIDPoint 500 blood gas analyser using blood gas syringes (BGSs) and to determine whether immediate analysis of venous blood collected into lithium heparin (LH) tubes can replace anaerobic blood sampling into BGSs. Methods Venous blood was collected from 24 healthy cats and directly transferred into a BGS and an LH tube. The BGS was immediately analysed on the RAPIDPoint 500 followed by the LH tube. The BGSs and LH tubes were compared using paired t-test or Wilcoxon matched-pairs signed-rank test, Bland-Altman and Passing-Bablok analysis. To assess clinical relevance, bias or percentage bias between BGSs and LH tubes was compared with the allowable total error (TEa) recommended for the respective parameter. Results Based on the values obtained from the BGSs, RIs were calculated for the evaluated parameters, including blood gases, electrolytes, glucose and lactate. Values derived from LH tubes showed no significant difference for standard bicarbonate, whole blood base excess, haematocrit, total haemoglobin, sodium, potassium, chloride, glucose and lactate, while pH, partial pressure of carbon dioxide and oxygen, actual bicarbonate, extracellular base excess, ionised calcium and anion gap were significantly different to the samples collected in BGSs ( P <0.05). Furthermore, pH, partial pressure of carbon dioxide and oxygen, extracellular base excess, ionised calcium and anion gap exceeded the recommended TEa. Conclusions and relevance Assessment of actual and standard bicarbonate, whole blood base excess, haematocrit, total haemoglobin, sodium, potassium, chloride, glucose and lactate can be made based on blood collected in LH tubes and analysed within 5 mins. For pH, partial pressure of carbon dioxide and oxygen, extracellular base excess, anion gap and ionised calcium the clinically relevant alterations have to be considered if analysed in LH tubes.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。