Experimental studies on environmental contamination with infected blood during haemodialysis

血液透析过程中受感染血液对环境污染的实验研究

阅读:2

Abstract

To assess the relative importance of different postulated modes of spread of hepatitis B in dialysis units, blood charged with various tracer organisms was used in simulated haemodialysis runs in four laboratories, and the resulting contamination of equipment and environment was measured semi-quantitatively. Some airborne spread of the tracer organism occurred when tubing containing contaminated blood was needled as the "patient" went on and came off the dialyser. Virtually no small airborne particles could be demonstrated however in simulated emergencies in which a blood line was disconnected, or even when bottles of blood were dropped on to a hard floor from a height of 2 metres. Bacillus globigii spores from contaminated blood leaked in small numbers into the dialysing fluid through apparently intact coils. T3 phage, with a particle size of the same order as hepatitis B virus, passed in small quantities through the membrane of a Kiil dialyser from blood to dialysing fluid and also in the reverse direction when added to the header tank. A number of other dialysers were also permeable to phage. Visual assessment of the appropriate moment for inserting the venous line into the "patient" at the onset of dialysis was shown to be unreliable, as the displaced fluid from the end of the venous line was already contaminated before it contained visible red blood cells. Considerable contamination of exposed surfaces and of the buttons on the proportionating unit cabinet occurred. Minor visible splashing of blood was a common-place of the laboratory experiments and was shown to be also a common event during routine haemodialysis in two of the dialysis units taking part in the studies.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。