Nondestructive Measurement of Hemoglobin in Blood Bags Based on Multi-Pathlength VIS-NIR Spectroscopy

基于多光程可见-近红外光谱的血袋中血红蛋白无损测量

阅读:1

Abstract

Hemoglobin concentration is an indicator for assessing blood product quality. To measure hemoglobin concentration in blood products without damaging blood bags, we proposed a method based on visible-near infrared transmission spectroscopy. Complex optical properties of blood bag walls result in measurement irregularities. Analyses showed that the slope of the light intensity-pathlength curve was more robust to the influence of the blood bag wall. In this study, the transmission spectra of red blood cell suspensions at multiple optical pathlengths were obtained, and the slopes of logarithmic light intensity-pathlength curves were calculated through curve fitting. A nondestructive measurement of hemoglobin content was achieved by using a regression model correlating slope spectra and hemoglobin concentration. Sixty samples with hemoglobin concentrations ranging from 72 to 161 g/L were prepared. Among them, 40 samples were used as a calibration set, and the remaining 20 samples were used as a prediction set. The determination coefficient of the prediction set was 0.97, with a mean square error of 2.78 g/L. This result demonstrates that a non-destructive measurement of hemoglobin levels in blood bags can be achieved by multiple-pathlength transmission spectroscopy.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。