Multicenter Study of Needle-Free Blood Collection System for Reducing Specimen Error and Intravenous Catheter Replacement

减少标本采集错误和静脉导管更换的无针采血系统多中心研究

阅读:1

Abstract

Inpatient hospital settings require access to high-quality blood specimens and durable peripheral intravenous (IV) catheters for patient care. The most common standard-of-care method for acquiring each blood specimen-venipuncture-often results in a non-negligible preanalytical error rate, patient discomfort, and tissue inflammation. In a 2-year, multicenter (23 hospitals) retrospective study, a novel blood collection system that collects blood specimens through existing inpatient IVs without a needle (PIVO, Velano Vascular) was compared with the current standard of care, in regards to its effect on specimen quality and IV catheter longevity. Using the PIVOTM device for blood collection decreased the rate of preanalytical errors by 56% compared with other collection methods, including venipuncture and conventional line draws. In addition, peripheral IV catheters that were used with PIVOTM for blood draws also had a 19% lower rate of replacement compared with those that did not. This is the largest study to date of PIVO use and demonstrates significant quality improvement outcomes compared with the current standard of care in blood collection, providing an opportunity to innovate inpatient hospital care.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。