Diagnostic accuracy of a bovine specific electronic beta-hydroxybutyrate handheld meter in fresh blood and stored serum samples

牛特异性电子β-羟基丁酸手持式检测仪在新鲜血液和储存血清样本中的诊断准确性

阅读:1

Abstract

This study aims to evaluate the diagnostic accuracy of the bovine specific beta-hydroxybutyrate (BHB) meter Nova Vet (NVET). We evaluated the accuracy and agreement of the NVET in fresh blood and thawed serum with the reference laboratory assay; and the repeatability, the interference by anticoagulants, and the optimum slope calibration factor. Individual blood samples were collected from 200 Holstein and crossbred cows between 3 and 14 days post-calving from 13 dairy herds in Minnesota. Using a laboratory assay with a cut point of 1.2 BHB mmol/L hyperketonemia prevalence was 10.6% (95% CI: 6.7, 15.8). The sensitivity of NVET in blood and serum was 100.0% while the specificity was 98.3 and 97.7% respectively. The agreement between NVET and the laboratory assay was the highest using blood samples (concordance correlation coefficient -CCC = 96.2, 95% CI: 95.0, 97.1. The coefficient of variation including within day (intra-meter), between- days, and -batches was 13.4% when testing blood samples. Minimal interference was observed with the use of anticoagulants (K-EDTA and Li-Heparin, CCC 0.90 and 0.93 respectively) in reference to whole blood without anticoagulant. The best calibration slope factor in serum was 1.0 (Youden's index: 0.98). Results suggest that the NVET device maintained a high accuracy and precision to quantified BHB concentration when applied in fresh blood and thawed serum under field conditions using the default calibration slope (1.0), and with minimal anticoagulant interference when used in whole blood samples.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。