Dynamic blood compatibility assessment of coronary stents using an innovative pneumatic blood pump to generate physiological flow rates in small sample chambers

利用创新的气动血泵在小型样本室中产生生理流速,对冠状动脉支架进行动态血液相容性评估

阅读:1

Abstract

According to ISO 10993-4, shear-dependent processes must be considered in the blood compatibility testing of cardiovascular implants, which demands controlled and physiological flow conditions in the test environment. For very small test objects such as coronary stents with a diameter below 5 mm, this is difficult to achieve, as previous test methods are either unable to represent the required flow velocities in small-lumen sample chambers, or can only do so under high pump shear stress, which in extreme cases can mask the effects of the tiny test objects. In this paper, we present a novel concept for dynamic in vitro models based on pneumatically generated blood flow (AirDrive technology), which can achieve flow rates and velocities up to 450 ml min(-1) or 930 mm s(-1), respectively. This allows for hemocompatibility testing of coronary stents under physiologically high flow rates without generating major shear stress from the pump mechanism. In an initial feasibility study, single coronary stents with different surface finishes (polished/unpolished) were subjected to flow velocities of up to 560 mm s(-1) in a 3.2 mm diameter sample chamber. Blood samples were collected before and after perfusion, and haematological and coagulation markers were analysed. Unpolished stents elicited higher shear-induced blood responses than polished stents. This demonstrates that our experimental setup is highly sensitive and enables precise and robust investigation of blood compatibility under physiologically relevant flow conditions, even for the smallest objects under investigation. To show the superiority of this novel model in creating high flow rates while maintaining minimal blood damage, the AirDrive system was compared to the commonly used roller pump closed-loop system, with the former exhibiting significantly less blood damage. This further confirms that the AirDrive technology we present in this paper is of the highest value for developmental or regulatory testing of blood compatibility of small- and medium-sized test objects.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。