Comparative analysis of image guidance in two institutions for prostate cancer patients

两家机构前列腺癌患者影像引导治疗的比较分析

阅读:3

Abstract

AIM/BACKGROUND: The analysis of systematic and random errors obtained from the pooled data on inter-fraction prostate motion during radiation therapy in two institutions. MATERIALS AND METHODS: Data of 6085 observations for 216 prostate cancer patients treated on tomotherapy units in two institutions of position correction shifts obtained by co-registration of planning and daily CT studies were investigated. Three independent variables: patient position (supine or prone), target (prostate or prostate bed), and imaging mode (normal or coarse) were analyzed. Systematic and random errors were evaluated and used to calculate the margins for different options of referencing based on the position corrections observed with one, three, or five imaging sessions. RESULTS: Statistical analysis showed that only the difference between normal and coarse modes of imaging was significant, which allowed to merge the supine and prone position sub-groups as well as the prostate and prostate bed patients. In the normal and coarse imaging groups, the margins calculated using systematic and random errors in the medio-lateral and cranio-caudal directions (5.5 mm and 4.5 mm, respectively) were similar, but significantly different (5.3 mm for the normal mode and 7.1 mm for the coarse mode) in the anterio-posterior direction. The reference scheme based on the first three fractions (R3) was found to be the optimal one. CONCLUSIONS: The R3 reference scheme effectively reduced systematic and random errors. Larger margins in the anterio-posterior direction should be used during prostate treatment on the tomotherapy unit, as coarse imaging mode is chosen in order to reduce imaging time and dose.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。