Establishment and application of autoverification system for HbA1c testing

建立和应用糖化血红蛋白(HbA1c)检测自动验证系统

阅读:3

Abstract

INTRODUCTION: This study aimed to determine autoverification rules for routine glycated hemoglobin (HbA1c) analysis based on high-performance liquid chromatography (HPLC) principle. Laboratory information system (LIS) and Bio-Rad D-100 Advisor software (Bio-Rad, Hercules, USA) with graphics recognition function were carriers for the autoverification system. MATERIALS AND METHODS: A total of 105,126 HbA1c results, including 98,249 HbA1c matching fast plasma glucose (FPG) results of real-world data from May 2019 to June 2020, were collected to determine autoverification rules including flags, delta checks, reporting limits, and logical rules. The validation database was composed of 48,045 HbA1c results and 41,083 matching FPG results. Autoverification passing rate and the reduction of turnaround time (TAT) were evaluated. RESULTS: Four autoverification systems (A, B, C, D) were established by two types of delta check rules, 28 flags, one reporting limits, and two kinds of logical rules. The autoverification passing rates were 80.6%, 78.8%, 83.7%, and 81.3%, and the average time saved in TAT were 117.5 min, 116.7 min, 121.1 min, and 121.7 min, respectively. CONCLUSIONS: Autoverification system C was the optimal one. Application of distribution of FPG corresponding to HbA1c groups had better performance as logical rules. Established HbA1c autoverifcation system shortened the auditing report time and improved work efficiency.