Evaluation of System Accuracy, Precision, Hematocrit Influence, and User Performance of Two Blood Glucose Monitoring Systems Based on ISO 15197:2013/EN ISO 15197:2015

基于 ISO 15197:2013/EN ISO 15197:2015 标准的两种血糖监测系统的系统准确度、精密度、血细胞比容影响和用户性能评估

阅读:1

Abstract

INTRODUCTION: Sufficiently high analytical quality of blood glucose monitoring systems (BGMS) is a prerequisite for efficient diabetes therapy. In this study we assessed system accuracy, measurement repeatability, intermediate measurement precision, user performance, and the influence of hematocrit on two CE-marked blood glucose monitoring systems. For one BGMS, measurement accuracy using venous samples was additionally investigated. METHODS: Study procedures were based on the International Organization of Standardization (ISO) 15197:2013/EN ISO 15197:2015 ("ISO 15197"). User performance included data from 100 subjects who used one test strip lot, whereas for all other analyses three different reagent system lots were used. For system accuracy assessment, 100 capillary samples were measured in duplicate with each of three reagent system lots per system, resulting in 600 results per system. RESULTS: CareSens S Fit and CareSens H Beat both fulfilled the ISO 15197 accuracy criteria with 97.5-100% of each test strip lot's results falling within ± 15 mg/dL or ± 15% of the results of the comparison method and 100% of results in consensus error grid (CEG) zone A for all three lots. User performance evaluation revealed sufficient accuracy in the hands of lay users although some handling errors were documented by study staff. Assessment of measurement repeatability and intermediate measurement precision is given by standard deviation (SD) (glucose levels < 100 mg/dL) and by coefficient of variation (CV) (glucose concentrations ≥ 100 mg/dL). SD was ≤ 4.1 mg/dL and CV ≤ 4.2% for measurement repeatability and SD was ≤ 2.2 mg/dL and CV ≤ 2.6% for intermediate measurement precision. In case of hematocrit influence, both BGMS complied with all three tested lots with the defined criteria. CONCLUSION: Both BGMS analyzed in this study fulfilled the required accuracy criteria of ISO 15197. They showed high precision, good performance in the hands of lay users, and the influence of hematocrit was acceptable in the labeled range.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。