Patch Pumps: Periodic Insulin Delivery Patterns

贴片式胰岛素泵:周期性胰岛素输注模式

阅读:1

Abstract

Recent in vitro experiments with patch pumps (PP) Omnipod (OP), Omnipod DASH (OP-D), A6 TouchCare (A6), and Accu-Chek Solo (ACS) have observed periodic fluctuations in the delivered amount of insulin during basal rate and consecutive bolus delivery in some PP, calling for a more systematic characterization of these periodic delivery patterns. Here, it was found that during basal rate delivery of 1 U/h, some devices of OP, OP-D, and A6 showed deviations of up to ±30% from target delivery that consistently repeated every 5 hours, whereas ACS showed no clear periodicity with considerably lower deviations. Similar results were found during consecutive bolus delivery of 1 U, where deviations repeated consistently every five boluses in some devices of OP, OP-D, and A6. However, there was a large variability in the periodic delivery patterns between individual devices of the same PP model. Examining their pumping techniques indicated a connection between the insulin delivery mechanism and observed delivery patterns of the PP. However, the clinical impact of such patterns is unclear.