Dose accuracy and injection force of disposable pens delivering pramlintide for the treatment of diabetes

用于治疗糖尿病的普兰林肽一次性注射笔的剂量准确性和注射力

阅读:1

Abstract

BACKGROUND: The pen injection format, typically used for insulin administration, has been adapted for the injectable, noninsulin diabetes therapy pramlintide. Administered before major meals, pramlintide therapy requires two to four injections/day in addition to the patients' usual insulin injections. The dose accuracy and injection force was determined for the 60 and 120 µg pramlintide pens. METHODS: Dose accuracy testing was conducted at two sites on multiple 60 µg (15, 30, and 60 µg doses) and 120 µg pens (60 and 120 µg doses) at prespecified temperatures (5-40 °C) and humidities (0-75%) using 29 G half-inch needles. All pens were stabilized under testing conditions for 4 h prior to testing. One site used a compression load cell (Zwick device) to test pens; one site performed tests manually. Injection-force testing was conducted at one site on multiple 60 and 120 µg pens at multiple temperatures (18-28 °C) and humidities (25-75%) using 29 and 31 G half-inch needles and an injection speed of 150 m/min. Injection-force testing was performed using a Zwick device. RESULTS: Dose accuracy for all pens tested, regardless of location, reproducibly met/exceeded acceptance criteria. Mean percentage of dose accuracy was 96.04 to 102.45% [standard deviations (SDs) 0.3 to 1.4 µg] for the 60 µg pen and 98.16 to 101.83% (SDs 0.4 to 2.5 µg) for the 120 µg pen. The average injection force across both pens did not exceed 7 N regardless of needle size. CONCLUSIONS: High dose accuracy and low injection force were observed for the 60 and 120 µg pens under a variety of conditions.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。