Optimization of Traction Magnetic Resonance Imaging to Improve Visibility of the Elbow Cartilage

优化牵引磁共振成像以提高肘关节软骨的可见度

阅读:1

Abstract

We previously reported that elbow magnetic resonance imaging (MRI) with 7 kg traction increases the joint space width of the radiocapitellar joint and improves articular cartilage visibility without arthrography. However, the optimal traction weight remains unclear. We assessed the effects of different traction weights on elbow MRI in 30 healthy volunteers. Elbow MRI was performed without traction and with 3, 5, and 7 kg axial tractions. The joint space width, humeral articular cartilage outline visibility, and intraprocedural pain/discomfort were evaluated. The joint and cartilage parameters were measured at the radiocapitellar joint and the lateral and medial thirds of the ulnohumeral joint. At the radiocapitellar joint, the joint space width increased significantly with traction. The cartilage outline visibility significantly increased with traction, with no significant differences among the traction weights. No significant result was observed at the lateral and medial thirds of the ulnohumeral joint. Pain and discomfort significantly increased as we used heavier traction weights. Elbow MRI with 3 kg traction showed sufficient effects similar to those observed with 7 kg traction with minimal pain and discomfort. There was no difference in the effect of traction between male and female participants. This procedure may enable enhanced visualization of intra-articular elbow injuries.