Tonometry in corneal edema after cataract surgery: dynamic contour tonometry versus Goldmann applanation tonometry

白内障手术后角膜水肿的眼压测量:动态轮廓眼压计与Goldmann压平眼压计的比较

阅读:1

Abstract

INTRODUCTION: Intraocular pressure (IOP) determination using dynamic contour tonometry (DCT) has been considered to be independent of central corneal thickness (CCT), while Goldmann applanation tonometry (GAT) is known to be influenced by various corneal properties. In this study, IOP was measured before and 1 day after cataract surgery using GAT and DCT to investigate the possible effects of corneal edema on IOP measurements. METHODS: Thirty patients with advanced cataracts were included in a pilot study. IOP was measured using GAT and DCT before and 1 day after phacoemulsification. CCT was determined before and after surgery to quantify postsurgical corneal edema. RESULTS: CCT increased significantly (by 89.7 ± 107.4 μm, P < 0.0001) 1 day after surgery. No significant difference was found for IOP measurements using GAT and DCT before surgery (mean IOP GAT: 17.5 ± 5.7 mmHg; mean IOP DCT: 17.9 ± 6.4 mmHg; P = 0.67) and 1 day after surgery (mean IOP GAT: 16.1 ± 6.6 mmHg; mean IOP DCT: 16.8 ± 8.3 mmHg; P = 0.69). IOP values using GAT and DCT were significantly correlated before as well as 1 day after surgery (before surgery: r = 0.82, P < 0.0001; after surgery r = 0.83, P < 0.0001). Bland-Altman plots showed a high variability in the difference in IOP measurements between methods before and 1 day after surgery. CONCLUSION: GAT and DCT seem to be equally valuable in IOP determination in postsurgical central corneal edema, although large differences between both methods are present in individual patients. IOP evaluation in corneal edema remains a difficult clinical challenge.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。