Intra- and interobserver reliability of ocular surface analyzer LacryDiag(®)

眼表分析仪 LacryDiag(®) 的观察者内和观察者间可靠性

阅读:1

Abstract

AIM: To invastigate intra- and interobserver reliability of interferometry, tear meniscus height (TMH) measurement and meibography (MBG) of an ocular surface analyzer, LacryDiag (Quantel Medical, France). METHODS: Five consecutive measurements and subsequent analysis of interferometry, TMH, and MBG were recorded by two examiners using the LacryDiag. To assess intra- and interobserver reliability, we used Cohen's kappa for categorical variables (interferometry), or intraclass correlation coefficient for continuous variables (TMH, MBG). RESULTS: Thirty eyes of 30 examinees were included. For both observers, there was excellent intraobserver reliability for MBG (0.955 and 0.970 for observer 1 and 2, respectively). Intraobserver reliability for observer 1 was substantial for interferometry (0.799), and excellent for TMH (0.863). Reliability for observer 2 was moderate for interferometry (0.535) and fair to good for TMH (0.431). Interobserver reliability was poor for interferometry (0.074) and fair to good for TMH (0.680) and MBG (0.414). CONCLUSION: LacryDiag ocular surface analyzer in our study proves to be a reliable noninvasive tool for the evaluation of TMH and MBG. As for interferometry, poor interobserver reliability, fair to good intraobserver reliability for observer 1, and moderate for observer 2, leave room for improvement.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。