Evaluation of Selected Biometric Parameters in Cataract Patients-A Comparison between Argos(®) and IOLMaster 700(®): Two Swept-Source Optical Coherence Tomography-Based Biometers

白内障患者特定生物测量参数的评估——Argos® 和 IOLMaster 700® 的比较:两种基于扫频源光学相干断层扫描的生物测量仪

阅读:2

Abstract

Background and Objectives: To compare the biometry of eyes obtained with two swept-source optical coherence tomography-based biometers-Argos (A), using an individual refractive index, and IOLMaster 700 (IM), using an equivalent refractive index-for all structures. Materials and Methods: The biometry of 105 eyes of 105 patients before cataracts were analyzed in this study. Parameters such as axial length (AL), anterior chamber depth (ACD), and lens thickness (LT) were compared from both devices. According to the axial length measurements, patients were divided into three groups, as follows: group 1-short eyes (AL < 22.5 mm), group 2-average eyes (22.5 ≤ AL ≤ 26.0 mm), and group 3-long eyes (AL > 26.0 mm). Results: The correlation coefficiency among all compared parameters varies from R = 0.92 to R = 1.00, indicating excellent reliability of IM and A. A statistical significance in axial length was indicated in the group of short eyes (n = 26)-mean AL (A) 21.90 mm (±0.59 mm) vs. AL (IM) 21.8 mm ± (0.61 mm) (p < 0.001)-and in the group of long eyes (n = 5)-mean AL (A) 27.95 mm (±2.62 mm) vs. mean AL (IM) 28.10 mm (±2.64) (p < 0.05). In the group of average eyes (n = 74), outcomes were similar-mean AL (A) 23.56 mm (±0.70 mm) vs. mean AL (IM) 23,56 mm (±0.71 mm) (p > 0.05). The anterior chamber depth measurements were higher when obtained with Argos than with IOLMaster 700-mean ACD (A) 3.06 mm (±0.48 mm) vs. mean ACD (IM) 2.92 mm (±0.46) p < 0.001. There was no statistical significance in mean LT-mean LT (A) 4.75 mm (±0.46 mm) vs. mean LT (IM) 4.72 mm (±0.44 mm) (p = 0.054). The biometry of one eye with dense cataracts could be measured only with Argos, using the Enhanced Retinal Visualization mode. Conclusions: Axial length measurements from both devices were different in the groups of short and long eyes, but were comparable in the group of average eyes. The anterior chamber depth values obtained with Argos were higher than the measurements acquired with IOLMaster 700. These differences may be particularly important when selecting IOLs for patients with extreme AL values.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。