Design and validation of a custom-made system to measure transepithelial electrical impedance in human corneas preserved in active storage machine

设计并验证了一种定制系统,用于测量保存在活性储存机中的人类角膜的跨上皮电阻抗。

阅读:1

Abstract

Corneal epithelial barrier represents one of the major limitations to ocular drug delivery and can be explored non-invasively through the evaluation of its electrical properties. Human corneas stored in active storage machine (ASM) could represent an interesting physiological model to explore transcorneal drug penetration. We designed a new system adapted to human corneas preserved in ASM to explore corneal epithelial barrier function ex-vivo. A bipolar set-up including Ag/AgCl electrodes adaptors to fit the corneal ASM and a dedicated software was designed and tested on freshly excised porcine corneas (n = 59) and human corneas stored 14 days in ASM (n = 6). Porcine corneas presented significant and proportional decrease in corneal impedance in response to increasing-size epithelial ulcerations and acute exposure to benzalkonium chloride (BAC) 0.01 and 0.05%. Human corneas stored 14 days in ASM presented a significant increase in corneal impedance associated with the restoration of a multi-layer epithelium and an enhanced expression of tight junctions markers zonula occludens 1, claudin 1 and occludin. These results support the relevance of the developed approach to pursue the exploration and development of human corneas stored in ASM as a physiological pharmacological model.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。