Gallium-68 DOTATATE Production with Automated PET Radiopharmaceutical Synthesis System: A Three Year Experience

使用自动 PET 放射性药物合成系统生产镓-68 DOTATATE:三年经验

阅读:6
作者:Alireza Aslani, Graeme M Snowdon, Dale L Bailey, Geoffrey P Schembri, Elizabeth A Bailey, Paul J Roach

Conclusion

In our experience the automated synthesis system performs reliably with a relatively low incident of failures. Our system had a consistent and reliable Ga-68 DOTATATE output with high labelling efficiency and purity. There is minimal operator intervention and radiation exposure. The system is GMP-compliant and has low maintenance and acceptable running costs. This system together with the recommended (68)Ge/(68)Ga generator is well suited for use in a hospital-based radiopharmacy.

Methods

The automated synthesis system we use is divided into three parts of a) servomotor modules, b) single use sterile synthesis cassettes and, c) a computerised system that runs the modules. An audit trail is produced by the system as a requirement for GMP production. The required reagents and chemicals are made in-. The Germanium breakthrough is determined on a weekly basis. Production yields for each synthesis are calculated to monitor the performance and efficiency of the synthesis. The quality of the final product is assessed after each synthesis by ITLC-SG and HPLC methods.

Results

A total of 500 Ga-68 DOTATATE syntheses (>800 patient doses) were performed between March 2011 and February 2014. The average generator yield was 81.3±0.2% for 2011, 76.7±0.4% for 2012 and 75.0±0.3% for 2013. Ga-68 DOTATATE yields for 2011, 2012, and 2013 were 81.8±0.4%, 82.2±0.4% and 87.9±0.4%, respectively. These exceed the manufacturer's expected value of approximately 70%. Germanium breakthrough averaged 8.6×10(-6)% of total activity which is well below the recommended level of 0.001%. The average ITLC-measured radiochemical purity was above 98.5% and the average HPLC-measured radiochemical purity was above 99.5%. Although there were some system failures during synthesis, there were only eight occasions where the patient scans needed to be rescheduled.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。