Assessment of the Accuracy of Ultrasonographically Measured Optic Nerve Sheath Diameter as a Surrogate for the Detection of Intracranial Hypertension Compared to Optic Nerve Sheath Diameter Measured by MRI: A Prospective Observational Study

超声测量视神经鞘直径作为颅内高压检测替代指标的准确性评估:与磁共振成像测量视神经鞘直径的比较:一项前瞻性观察研究

阅读:2

Abstract

BACKGROUND:  Ultrasonographic measurement of optic nerve sheath diameter (ONSD) has been proposed as a non-invasive, bedside method to detect raised intracranial pressure (ICP) in various clinical settings. We aimed to correlate the ONSD obtained by ultrasonography (USG) with the ONSD obtained by magnetic resonance imaging (MRI) and to find its measurement accuracy. METHODOLOGY:  A prospective double-blind study was carried out by performing ocular ultrasounds on 32 patients with clinical features of intracranial hypertension. ONSD was measured by both USG and MRI. Three ultrasonographic scans were performed in both the eyes and the average value of the same was correlated with ONSD obtained from MRI. The three ultrasonographic ONSD scans were also compared among each other for reproducibility and intra-observer variability. RESULTS:  There was a significant correlation between the ONSD measured by MRI and USG (r = 0.954, p = 0.0000, mean difference <5%). A notable degree of agreement between the two measurement modalities was found by the Bland-Altman test (coefficient = -0.116, p = 0.050). CONCLUSION:  ONSD measured by USG has a significant degree of agreement with that of MRI and hence can serve as an effective, fast, and reliable evaluation technique in patients with intracranial hypertension.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。