Development and Clinical Validation of a Sensitive Lateral Flow Assay for Rapid Urine Fentanyl Screening in the Emergency Department

用于急诊科快速尿液芬太尼筛查的灵敏侧流检测法的开发和临床验证

阅读:5
作者:Zhao Li, Hui Chen, Sheng Feng, Kengku Liu, Ping Wang

Background

Rapid identification of fentanyl at the point-of-care is critical. Urine fentanyl concentrations in overdose cases start at single-digit nanograms per milliliter. No fentanyl point-of-care assay with a cutoff at single-digit nanograms per milliliter is available.

Conclusions

The LFA can detect fentanyl and norfentanyl with high clinical sensitivity and specificity in the ED population with rapid fentanyl screening needs.

Methods

A competitive lateral flow assay (LFA) was developed using gold nanoparticles and optimized for rapid screening of fentanyl in 5 minutes. Urine samples from 2 cohorts of emergency department (ED) patients were tested using the LFA and LC-MS/MS. The 2 cohorts consisted of 218 consecutive ED patients with urine drug-of-abuse screen orders and 7 ED patients with clinically suspected fentanyl overdose, respectively.

Results

The LFA detected fentanyl (≥1 ng/mL) and the major metabolite norfentanyl (≥10 ng/mL) with high precision. There was no cross-reactivity with amphetamine, cocaine, morphine, tetrahydrocannabinol, methadone, buprenorphine, naloxone, and acetaminophen at 1000 ng/mL and 0.03%, 0.4%, and 0.05% cross-reactivity with carfentanil, risperidone, and 9-hydroxyrisperidone, respectively. In 218 consecutive ED patients, the prevalence of cases with fentanyl ≥1 ng/mL or norfentanyl ≥10 ng/mL was 5.5%. The clinical sensitivity and specificity of the LFA were 100% (95% CI, 75.8-100%) and 99.5% (95% CI, 97.3-99.9%), respectively. The positive and negative predictive values were 92.3% (95% CI, 66.7-98.6%) and 100% (95% CI, 98.2-100%), respectively. The concordance between the LFA and LC-MS/MS was 100% in the 7 suspected fentanyl overdose cases (5 positive, 2 negative). Conclusions: The LFA can detect fentanyl and norfentanyl with high clinical sensitivity and specificity in the ED population with rapid fentanyl screening needs.