Accuracy of intraoperative neuromonitoring during percutaneous cement discoplasty

经皮椎间盘成形术中神经监测的准确性

阅读:1

Abstract

INTRODUCTION: Percutaneous cement discoplasty is a minimally invasive procedure to treat low back pain due to advanced degenerative disc disease in elderly patients. Complications of this procedure has been described such as infection, vertebral fracture, cement leakage and nerve injury. Intraoperative neuromonitoring is used to detect the latter. The objective of this study was to assess the usefulness of neuromonitoring during discoplasty to detect new neurological compromise. METHODS: 100 consecutive patients were included in this retrospective study, (30 males and 70 females, mean age of 76.3 ± 5.71 years) with mechanical low back pain who underwent percutaneous cement discoplasty. RESULTS: Sensitivity to detect neurological injury was 82% (CI 95% 66-98), specificity was of 99% (CI 95%98-100) with a positive predictive value of 0.95 (CI 95% 85-100) and a negative predictive value of 0.97 (CI 95% 95-99). In 5 patients neurological compromise was not detected by neuromonitoring. DISCUSSION: Our study showed high sensitivity and specificity of neuromonitoring to detect neurological irritation during percutaneous discoplasty. Intraoperative neuromonitoring resulted an effective assistance during this minimally invasive procedure.