3D Printed Microfluidic Devices for Solid-Phase Extraction and On-Chip Fluorescent Labeling of Preterm Birth Risk Biomarkers

用于固相萃取和芯片上荧光标记早产风险生物标志物的3D打印微流控装置

阅读:2

Abstract

Solid-phase extraction (SPE) is a general preconcentration method for sample preparation that can be performed on a variety of specimens. The miniaturization of SPE within a 3D printed microfluidic device further allows for fast and simple extraction of analytes while also enabling integration of SPE with other sample preparation and separation methods. Here, we present the development and application of a reversed-phase lauryl methacrylate-based monolith, formed in 3D printed microfluidic devices, which can selectively retain peptides and proteins. The effectiveness of these SPE monoliths and 3D printed microfluidic devices was tested using a panel of nine preterm birth biomarkers of varying hydrophobicities and ranging in mass from 2 to 470 kDa. The biomarkers were selectively retained, fluorescently labeled, and eluted separately from the excess fluorescent label in 3D printed microfluidic systems. These are the first results demonstrating microfluidic analysis processes on a complete panel of preterm birth biomarkers, an important step toward developing a miniaturized, fully integrated analysis system.