Stability of intrinsic rhythm in pacemaker-dependent patients during pacemaker replacement: Can we predict the need for temporary pacing?

起搏器依赖型患者在起搏器更换期间固有节律的稳定性:我们能否预测临时起搏的需求?

阅读:1

Abstract

BACKGROUND: In pacemaker-dependent patients, the risk of asystole must be managed during device replacement. This study aimed to examine whether we could predict the indication for temporary pacing (TP) during the generator replacement. METHODS: We studied 105 consecutive patients who underwent pacemaker replacement due to battery depletion at Juntendo Nerima Hospital between September 2005 and December 2016. We examined the relationship between the stability of the intrinsic rhythm (IR) during pacemaker replacement and several clinical factors including age, gender, sick sinus syndrome or atrioventricular (AV) block, duration of pacing, structural heart disease, use of anti-arrhythmic drugs, and the presence/absence of a stable IR (>40 bpm) at the outpatient clinic (OPC) just before pacemaker replacement. RESULTS: Of the 105 patients, we excluded 1 patient who required TP because of bradycardia-dependent torsades de points. Therefore, we evaluated 104 patients for the indication for TP. TP was underwent in 19 patients (18%) because of an absence or instability of the IR during pacemaker replacement. The indication for TP was significantly correlated with AV block (84% vs 48%, P = .0044) and the absence of a stable IR at the last OPC visit (89% vs 24%, P < .0001). For predicting the indication for TP, the following values of no stable IR at the last OPC visit were obtained: 89% sensitivity, 77% specificity, 46% positive predictive value, and 97% negative predictive value. CONCLUSIONS: The presence of a stable IR at the last OPC visit was a good predictor (97%) of no indication for TP during pacemaker replacement.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。