Soft Tissue Attenuation Patterns Associated with Upright Acquisition SPECT Myocardial Perfusion Imaging: A Descriptive Study

直立位SPECT心肌灌注显像相关的软组织衰减模式:一项描述性研究

阅读:1

Abstract

BACKGROUND: Little is known about soft tissue attenuation artifacts when an upright patient-position SPECTmyocardial perfusion imaging (MPI) system is used. In this investigation we sought to describe the patterns and frequency of attenuation artifacts associated with this type of instruments and we explored the impact of gender and body habitus on these artifacts. METHODS: In a cross-sectional study, we described the prevalence of various soft-tissue attenuation patterns in 212 normal SPECT-MPI studies acquired with an upright patient-position imaging system. RESULTS: In these 212 normal, clinically-indicated, upright-acquisition SPECT-MPIs the attenuation patterns observed were: anterior (6.1%), inferior (63.7%) and lateral (24.1%). Though uncommon, anterior attenuation trended to being more prevalent among women [9.5% vs. 3.4%, P=0.07] and was independently associated with chest circumference. Lateral attenuation was more common among women [34.7% vs. 15.4%, p=0.001] and was strongly associated with obesity (p<0.001). Inferior attenuation was more prevalent among men than women (75.2% vs. 49.5% respectively, P<0.001). CONCLUSIONS: Soft-tissue attenuation artifacts are common in upright-acquisition SPECT-MPI. Recognizing the frequency of these attenuation patterns and their interaction with gender and body habitus is critical for the accurate interpretation of SPECT-MPI.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。