Influence of Milk on Exhaled Carbon Monoxide (CO) Measurement by Portable CO Monitors

牛奶对便携式一氧化碳监测仪测量呼出气一氧化碳(CO)的影响

阅读:1

Abstract

BACKGROUND: A portable breath carbon monoxide (CO) monitor has a high cross-sensitivity to hydrogen (H(2)). This study examined the influences of H(2) after consuming milk on the detected CO values using three types of portable CO monitors. MATERIALS AND METHODS: Exhaled breath from seven participants (four healthy nonsmokers and three smokers with otherwise unknown comorbidities) was collected in sampling bags. The participants then consumed 200 mL of milk, and the exhaled breath of each was collected in separate bags every 30 minutes until 9 hours later. CO and H(2) in the bag were measured using a gas chromatograph as a reference analyzer, and CO was also measured using three types of portable CO monitors. RESULTS: After consuming milk, H(2) levels were significantly higher, and CO levels were not significantly elevated as measured by the reference analyzer. However, CO levels in monitors A and B were significantly elevated, even though participants did not smoke. The H(2) levels in the reference analyzer significantly increased and reached a maximum 4.5 hours after consuming milk. The difference in CO levels between the reference analyzer and each monitor increased significantly after 5 or 5.5 hours. CONCLUSIONS: This study suggested that the breath CO monitors with a cross-sensitivity to H(2) responded to H(2) as CO in the exhaled gas and measured higher than actual values after milk consumption. The extent of the influence of H(2) differed depending on the type of CO monitor. It is necessary to consider milk consumption when assessing the smoking status of people using portable CO monitors.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。