Test characteristics of two rapid antigen detection tests (SD FK50 and SD FK60) for the diagnosis of malaria in returned travellers

两种快速抗原检测试剂盒(SD FK50 和 SD FK60)在诊断归国旅行者疟疾方面的检测特性

阅读:1

Abstract

BACKGROUND: Two malaria rapid diagnostic tests were evaluated in a travel clinic setting: the SD FK50 Malaria Ag Plasmodium falciparum test (a two-band test) and the SD FK60 Malaria Ag P. falciparum/Pan test (a three-band test). METHODS: A panel of stored whole blood samples (n = 452 and n = 614 for FK50 and FK60, respectively) from returned travellers was used. The reference method was microscopy with PCR in case of discordant results. RESULTS: For both tests, overall sensitivity for the detection of P. falciparum was 93.5%, reaching 97.6% and 100% at parasite densities above 100 and 1,000/microl respectively. Overall sensitivities for Plasmodium vivax, Plasmodium ovale and Plasmodium malariae for the FK60 test were 87.5%, 76.3% and 45.2%, but they reached 92.6% and 90.5% for P. vivax and P. ovale at parasite densities above 500/microl. Specificities were above 95% for all species and both tests when corrected by PCR, with visible histidine-rich protein-2 lines for P. malariae (n = 3) and P. vivax and P. ovale (1 sample each). Line intensities were reproducible and correlated to parasite densities. The FK60 tests provided clues to estimate parasite densities for P. falciparum below or above 1,000/microl. CONCLUSION: Both the FK50 and FK60 performed well for the diagnosis of P. falciparum in the present setting, and the FK60 for the diagnosis of P. vivax and P. ovale at parasite densities > 500/microl. The potential use of the FK60 as a semi-quantitative estimation of parasite density needs to be further explored.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。