Spatial resolution improvements for photon counting detectors using coincidence counting and frequency weighted reconstruction

利用符合计数和频率加权重建提高光子计数探测器的空间分辨率

阅读:1

Abstract

BACKGROUND: Photon counting X-ray detectors (PCDs) provide better spatial resolution than energy integrating X-ray detectors, but even higher resolution is desired in some applications. Certain charge sharing compensation techniques such as coincidence counting preferentially detect photons that arrive at the boundary between pixels, and this could be used for subpixel localization of incident photons and enhanced spatial resolution. PURPOSE: To estimate improvements to spatial resolution and detective quantum efficiency that are possible when using coincidence counting for high-resolution, non-spectral imaging. METHODS: The modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) were estimated using numerical simulations of a two-dimensional parallel-beam CT system. Coincidence counters were modeled either geometrically or with Monte Carlo simulations. The geometric model consisted of narrow coincidence counters that interlaced with wider ordinary pixels, and showed that using standard filtered backprojection decreases low-frequency DQE, but that a frequency weighting technique could be used to restore DQE. The Monte Carlo simulations were used to estimate the possible improvements that could be expected from real systems. The pixel pitch was 0.25 mm and the source apertures considered were 0, 0.125, and 0.5 mm. A numerical stent phantom was also used to illustrate possible improvements. RESULTS: Assuming a 0.125 mm source aperture and the Monte Carlo model, the limiting MTF (at 10%) increased from 20 to 40 lp/cm using coincidence counters. This can be explained by the increased sampling (and higher Nyquist limit) possible from coincidence counters. In the geometric model, coincidence counters are compared to conventional double sampling techniques such as in-plane flying focal spot, and the MTF at 38 lp/cm increased from 5% to 53%. Without the frequency weighting technique, low-frequency DQE was reduced by about 20%, but these losses are recovered with frequency weighting. Improvements are much more modest with the 0.25 mm source aperture because the system becomes source-limited. CONCLUSIONS: Coincidence counting could be used to increase spatial resolution in PCDs. The increases in system resolution could be large if a high-resolution X-ray source were available.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。