Skin-adaptive focused flexible micromachined ultrasound transducers for wearable cardiovascular health monitoring

用于可穿戴心血管健康监测的皮肤自适应聚焦柔性微机械超声换能器

阅读:2

Abstract

Continuous monitoring of cardiovascular vital signs can reduce the incidence and mortality of cardiovascular diseases, yet cannot be implemented by current technologies because of device bulkiness and rigidity. Here, we report self-adhesive and skin-conformal ultrasonic transducer arrays that enable wearable monitoring of multiple hemodynamic parameters without interfering with daily activities. A skin-adaptive focused ultrasound method with rational array design is proposed to implement measurement under wide ranges of skin curvatures and depths with improved sensing performances. We introduce a self-adhesive hydrogel layer to enhance the acoustic impedance matching, biocompatibility and contact reliability at device-skin interfaces, and a microelectromechanical systems-compatible process for batch-to-batch and high-consistency manufacturing. The resultant transducer array, with a frequency of 2.4 megacycles per second and a -6 dB bandwidth of 47%, implements accurate and continuous detection of cardiovascular signs, including blood pressure (BP) waveforms of various arteries, heart rate and BP during exercises, and arterial stiffness, validated clinically.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。