Development of LC-MS/MS Confirmatory Method for the Determination of Testosterone in Bovine Serum

建立 LC-MS/MS 确认方法测定牛血清中的睾酮

阅读:7
作者:Barbara Woźniak, Iwona Matraszek-Żuchowska, Sebastian Witek, Andrzej Posyniak

Conclusions

The developed method is specific and sensitive, suitable for measuring the natural level of testosterone in blood of cattle and for use in routine control programme for the detection of this hormone in bovine serum.

Material and methods

17β-testosterone and internal standards of 17%-testosterone-d2 were extracted from serum samples with a mixture of tert-butyl methyl ether/petroleum ether and were directly analysed by an LC/MS/MS on QTRAP 5500 instrument with a TurboIon-Spray source operating in a positive ionisation mode. Chromatographic separation was achieved on the analytical column Inertsil® ODS-3 with an isocratic elution using mobile phase consisting of acetonitrile, methanol, and water. Method validation has been carried out in accordance with the Commission Decision 2002/657/EC.

Methods

17β-testosterone and internal standards of 17%-testosterone-d2 were extracted from serum samples with a mixture of tert-butyl methyl ether/petroleum ether and were directly analysed by an LC/MS/MS on QTRAP 5500 instrument with a TurboIon-Spray source operating in a positive ionisation mode. Chromatographic separation was achieved on the analytical column Inertsil® ODS-3 with an isocratic elution using mobile phase consisting of acetonitrile, methanol, and water. Method validation has been carried out in accordance with the Commission Decision 2002/657/EC.

Results

The method was characterised by good recovery (82%) and precision (R.S.D 17 %). Decision limit (CCα) and detection capability (CCβ) was 0.05 μg L-1 and 0.09 μg L-1 respectively. The method met the criteria set out in Commission Decision 2002/657/EC for the purpose of confirmation in terms of retention time and ion ratio in the whole range of its application. Conclusions: The developed method is specific and sensitive, suitable for measuring the natural level of testosterone in blood of cattle and for use in routine control programme for the detection of this hormone in bovine serum.