Comparison of EV characterization by commercial high-sensitivity flow cytometers and a custom single-molecule flow cytometer

利用商用高灵敏度流式细胞仪和定制单分子流式细胞仪对细胞外囊泡进行表征的比较

阅读:8
作者:James Kim ,Shihan Xu ,Seung-Ryoung Jung ,Alya Nguyen ,Yuanhua Cheng ,Mengxia Zhao ,Bryant S Fujimoto ,Wyatt Nelson ,Perry Schiro ,Jeffrey L Franklin ,James N Higginbotham ,Robert J Coffey ,Min Shi ,Lucia N Vojtech ,Florian Hladik ,Muneesh Tewari ,John Tigges ,Ionita Ghiran ,Tijana Jovanovic-Talisman ,Louise C Laurent ,Saumya Das ,Olesia Gololobova ,Kenneth W Witwer ,Tuoye Xu ,Al Charest ,Kendall Van Keuren Jensen ,Robert L Raffai ,Jennifer C Jones ,Joshua A Welsh ,John P Nolan ,Daniel T Chiu

Abstract

High-sensitivity flow cytometers have been developed for multi-parameter characterization of single extracellular vesicles (EVs), but performance varies among instruments and calibration methods. Here we compare the characterization of identical (split) EV samples derived from human colorectal cancer (DiFi) cells by three high-sensitivity flow cytometers, two commercial instruments, CytoFLEX/CellStream, and a custom single-molecule flow cytometer (SMFC). DiFi EVs were stained with the membrane dye di-8-ANEPPS and with PE-conjugated anti-EGFR or anti-tetraspanin (CD9/CD63/CD81) antibodies for estimation of EV size and surface protein copy numbers. The limits of detection (LODs) for immunofluorescence and vesicle size based on calibration using cross-calibrated, hard-dyed beads were ∼10 PE/∼80 nm EV diameter for CytoFLEX and ∼10 PEs/∼67 nm for CellStream. For the SMFC, the LOD for immunofluorescence was 1 PE and ≤ 35 nm for size. The population of EVs detected by each system (di-8-ANEPPS+/PE+ particles) differed widely depending on the LOD of the system; for example, CellStream/CytoFLEX detected only 5.7% and 1.5% of the tetraspanin-labelled EVs detected by SMFC, respectively, and median EV diameter and antibody copy numbers were much larger for CellStream/CytoFLEX than for SMFC as measured and validated using super-resolution/single-molecule TIRF microscopy. To obtain a dataset representing a common EV population analysed by all three platforms, we filtered out SMFC and CellStream measurements for EVs below the CytoFLEX LODs as determined by bead calibration (10 PE/80 nm). The inter-platform agreement using this filtered dataset was significantly better than for the unfiltered dataset, but even better concordance between results was obtained by applying higher cutoffs (21 PE/120 nm) determined by threshold analysis using the SMFC data. The results demonstrate the impact of specifying LODs to define the EV population analysed on inter-instrument reproducibility in EV flow cytometry studies, and the utility of threshold analysis of SMFC data for providing semi-quantitative LOD values for other flow cytometers.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。