Accuracy of the HemoCue® to measure cell-free plasma hemoglobin and detect clinically significant hemolysis

HemoCue®测量游离血浆血红蛋白和检测具有临床意义的溶血的准确性

阅读:3

Abstract

IntroductionMonitoring cell-free plasma hemoglobin (PHb) during extracorporeal therapies allows early intervention of significant hemolysis, but timely measurements are often challenging. We thus present an analysis of a rapid benchtop device's ability to detect clinically significant hemolysis (PHb ≥50 mg/dL).MethodsPHb was measured in 419 plasma samples from 88 pediatric patients undergoing cardiopulmonary bypass via both the benchtop device (HemoCue® Plasma/Low Hb system) and the clinical laboratory at the Children's Hospital of Pittsburgh (reference standards). Values of PHb ≥50 mg/dL as measured by the reference standard was defined as the binary outcome of clinically significant hemolysis. Analyses included Pearson correlations, logistic regression, receiver operating characteristic curves, and Bland-Altman. Because the manufacturer specifications identify the measurement range of the HemoCue® system as 30-3000 mg/dL, a secondary analysis was completed using PHb ≥30 mg/dL.ResultsUsing reference measurements, 66/88 subjects had at least one PHb level that fell within the range of detection (≥30 mg/dL) of the benchtop device and 46/88 had significant hemolysis as defined by PHb ≥50 mg/dL. PHb levels ≥30 mg/dL largely correlated with measurements made with the benchtop device (r = 0.82, p < .001). The device was able to predict PHb values ≥30 mg/dL (AUROC 0.9582) and ≥50 mg/dL (AUROC 0.9637). The Bland-Altman demonstrated a mean difference of 7.0 mg/dL with <5% outside the 95% limits of agreement.ConclusionsThe HemoCue® system is an effective surrogate for more robust laboratory testing to identify clinically significant hemolysis during cardiopulmonary bypass.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。