Automated Global Positioning Layout (GPL) for accuracy assessment in CAD-CAM mandibular reconstruction - Method validation

用于CAD/CAM下颌骨重建精度评估的自动全球定位布局(GPL)——方法验证

阅读:1

Abstract

The lack of a standardized methodology complicates accuracy assessment in computer-assisted mandibular reconstruction. Existing landmark-based methods are susceptible to operator variability, while surface-based comparisons can mask local deviations. This study validates a novel, automated protocol, the Global Positioning Layout (GPL), to quantify the 3D discrepancy between the virtual surgical plan and the postoperative outcome, by comparing its performance and reliability against Methods A and B. A retrospective cohort of 17 patients was analysed, with three operators performing all measurements on two occasions. The GPL method demonstrated complete reproducibility, with no inter- or intra-operator variability, providing a detailed, spatially-oriented assessment of deviations. In contrast, the landmark-based method showed poor reproducibility and systematic bias and was often inapplicable due to the absence of landmarks after resection. The surface-based method, while objective for its mean error metric, was operator-dependent for initial alignment and its non-directional output masked significant localized deviations. This study validates GPL as a robust and fully reproducible tool that overcomes the critical limitations of established techniques. The GPL method provides a strong foundation for a standardized protocol, essential for the reliable comparison of surgical outcomes, refinement of surgical techniques, and improvement of long-term patient outcomes.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。