Commissioning of LUNA three-dimensional surface-guided radiotherapy in full rotating gantry proton therapy system

LUNA三维表面引导放射治疗系统在全旋转机架质子治疗系统中的调试

阅读:2

Abstract

BACKGROUND: Surface-guided radiotherapy (SGRT) is a real-time, non-ionizing optical imaging technology that monitors patient surfaces during radiation treatment by creating 3D reconstructions from surface coordinate measurements. LAP has recently released a new SGRT solution, and we are the first proton therapy center to adopt this SGRT solution in a full rotating gantry. This work aims to present the commissioning results and quality assurance procedures at our center and to highlight the challenges of installing SGRT on a full rotating gantry. METHODS: Following closely the recommendations in TG-302, 6 different measurements were performed during the commissioning of the SGRT system, which was mounted linearly on the same beam due to space constraints. These measurements included spatial drift measurement, static localization accuracy, dynamic localization accuracy, end-to-end localization accuracy, field-of-view (FOV) characterization of different treatment sites, and stability upon gantry rotation. These measurements subsequently served as baseline values for the different quality assurance procedures recommended in TG-302. RESULTS: The moving average for spatial drift measurement across 200 minutes was 0.0233 mm, while the average deviation for static and dynamic localization was less than 0.2 mm across. The largest deviation was 0.182 mm in the LR direction at 40 mm for static localization, while dynamic localization has a maximum deviation from the Anzai laser tracking at 0.1 minute point with a value of 0.734 mm. End-to-end localization test reported a difference of 0.08 cm between x-ray imaging and LUNA 3D's isocenter. The FOV was in general mostly visible for all treatment sites except for the lower neck region and lateral prostate surfaces due to the placement of the cameras. Stability during gantry rotation is kept within 0.2 mm with gantry angle 30.4° and 335.8° having the largest magnitude deviations of 0.151 and 0.156 mm, respectively. CONCLUSION: The successful commissioning of the LUNA 3D SGRT system at our center can serve as a reference for other particle therapy centers with a full rotating gantry. This work will allow them to include the relevant tests as well as understand the limitations of FOV for different treatment sites when adopting similar camera arrangement like ours in a full rotating gantry.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。