Improved porous silicon microarray based prostate specific antigen immunoassay by optimized surface density of the capture antibody

通过优化捕获抗体的表面密度改进基于多孔硅微阵列的前列腺特异性抗原免疫测定

阅读:8
作者:SangWook Lee, Soyoun Kim, Johan Malm, Ok Chan Jeong, Hans Lilja, Thomas Laurell

Abstract

Enriching the surface density of immobilized capture antibodies enhances the detection signal of antibody sandwich microarrays. In this study, we improved the detection sensitivity of our previously developed P-Si (porous silicon) antibody microarray by optimizing concentrations of the capturing antibody. We investigated immunoassays using a P-Si microarray at three different capture antibody (PSA - prostate specific antigen) concentrations, analyzing the influence of the antibody density on the assay detection sensitivity. The LOD (limit of detection) for PSA was 2.5 ng mL(-1), 80 pg mL(-1), and 800 fg mL(-1) when arraying the PSA antibody, H117 at the concentration 15 μg mL(-1), 35 μg mL(-1), and 154 μg mL(-1), respectively. We further investigated PSA spiked into human female serum in the range of 800 fg mL(-1) to 500 ng mL(-1). The microarray showed a LOD of 800 fg mL(-1) and a dynamic range of 800 fg mL(-1) to 80 ng mL(-1) in serum spiked samples.